Owner's manual
<7. Startup>
7-14
IM 11M12A01-04E
7.9.3 Checking Input Contacts
Follow Table 7.10 to check the input contacts. The table uses an example with input contact 1.
Table 7.10 Checking Input Contacts
Switch operation Display Description
> ∧
ENT
A01
Display after the password has been entered.
> ∧
ENT
G01
Touch the [∧] key to switch to Group G.
> ∧
ENT
G01
Touch the [>] key to move the position of the digit that is ashing to the right one
digit.
> ∧
ENT
G21
Touch the [∧] key to enter 2.
> ∧
ENT
0
Touch the [ENT] key. 0 is displayed with the contact open. If the contact is closed,
the display will be 1 (one).
This enables you to check whether or not the wiring connections have been
properly made or not.
> ∧
ENT
G21
Touch the [ENT] key once again to switch to the parameter code selection display.
> ∧
ENT Basic
panel
display
Touch the [>] key together with the [ENT] key to return to the basic panel display.
The symbol [ ] indicates that the key is being touched. Light characters indicates that the digits are ashing.