Owner's manual

<7. Startup>
7-5
IM 11M12A01-04E
Table 7.2 Display Functions
Display Function and item to be set
Basic panel Displays the oxygen concentration in normal operation, or displays the detector
heater temperature while warming up. If an error or alarm arises,
the corresponding error or alarm number appears.
Password entry Enters the password for the parameter code selection display.
Group A setup Displays detailed data, such as the cell voltage or temperature.
Group B setup Sets and performs calibration.
Group C setup Sets analog output.
Group D setup Sets an alarm.
Group E setup Sets the input and output contacts.
Group F setup Selects the type of equipment and sets the parameters for computation.
Group G setup Performs the current-loop or contact checks.
7.4.3 Entering Parameter Code Selection Display
This section briey describes the password entry procedure for entering the parameter code selection
display. The password is 1102 - it cannot be changed to a different password.
Switch operation Display Description
>
ENT
21.0%
Warm-up is complete, and the basic panel is now displayed.
>
ENT
PASSno
Continuously touch the [ENT] key for at least three seconds
to display "PASSno."
>
ENT
0000
Touch the [ENT] key again. This allows you to change the leftmost
digit that is ashing.
>
ENT
1000
Set the password 1102. If you touch the [] key, the digit that is ashing will
be 1.
>
ENT
1000
Touch the [>] key to move the position of the digit that is ashing to the
right one digit.
>
ENT
1100
Touch the [] key to change the numeric value to 1.
>
ENT
1100
Touch the [>] key again to move the position of the digit that is ashing to
the right one more digit. Continuously touch the [>] key, and the position of
the digit that is ashing will move continuously to the right.
>
ENT
1102
Touch the [] key to change the numeric value to 2. Continuously touch []
key, and the numeric value increases continuously.
>
ENT
1102
If you touch the [ENT] key, all the digits ash.
>
ENT
A01
Touch the [ENT] key again to display A01 on the parameter code selection
display.
The symbol [ ] indicates that the key is being touched. Light characters indicate that the digits are ashing.
CAUTION
If no key is touched for at least 20 seconds during password entry, the current display will
automatically switch to the basic panel display.
If no key is touched for at least 10 minutes during parameter code selection, the current display
will automatically switch to the basic panel display.