5.5.2 Wiring Procedure
(1) M4 screws are used for the terminal of the converter. Each cable should be equipped with the
corresponding crimp contact.
(2) The ON/OFF level of this contact input is identied by the resistance. Connect a contact input
that satises the descriptions in Table 5.2.
Table 5.2 Identication of Contact Input ON/OFF
Closed Open
Resistance 200 Ω or less 100 kΩ or more