Owner's manual

<3. Installation>
3-6
IM 11M12A01-04E
3.2.2 Mounting of ZA8F Flow Setting Unit
The ow setting unit can be mounted either on a pipe (nominal JIS 50 A) or on a wall. It should be
positioned vertically so that the owmeter works correctly.
<Pipe Mounting>
(1) Prepare a vertical pipe of sufcient strength (nominal JIS 50A: O.D. 60.5 mm) for mounting the
ow setting unit. (The unit weighs approximately 2 to 3.5 kg.)
(2) Mount the ow setting unit on the pipe by tightening the nuts with the U-bolt so that the metal
tting is rmly attached to the pipe.
F3401E.ai
Figure 3.12 Pipe Mounting
<Wall Mounting>
(1) Make a hole in the wall as illustrated in Figure 3.13.
223
140
F3.13E.ai
4 - Ø6 hole, or M5 screw
Unit: mm
Figure 3.13 Mounting holes