Owner's manual
<10. Other Functions>
10-10
IM 11M12A01-04E
Table 10.6 Resetting
Switch operation Display Brief Description
> ∧
ENT
Err-01
/------
If an error occurs, the error number and "------" are displayed alternately,
as given on the left.
> ∧
ENT
PASSno
Hold down the [ENT] key for at least three seconds.
> ∧
ENT
0000
Touch the [ENT] key again to switch to the password entry display.
> ∧
ENT
1000
Enter the password 1102.
Intermediate switch operations omitted.
> ∧
ENT
1102
> ∧
ENT
A01
> ∧
ENT
G01
Change the parameter code to "G30".
> ∧
ENT
G01
> ∧
ENT
G30
> ∧
ENT
All the digits light
up.
Touch the [ENT] key to execute resetting.
The symbol [ ] indicates that the corresponding keys are being touched, and the light characters indicate “ ashing.”
“ / ” indicates that the characters are displayed alternately.