Owner manual
IM 11M12A01-05E
8-1
8. Detailed Data Setting
8. Detailed Data Setting
8.1 Current Output Setting
This section describes setting of the analog output range. Table 8.1 shows the parameter
codes for each setting item and set values.
Table 8.1 Display Items
T8.1.1.EPS
Set item Parameter code Set value
Current output C01 0: Oxygen concentration
1: Humidity
2: Mixing ratio
Output mode C03 0: Linear
1: Logarithm
Min. oxygen concentration C11 Oxygen concentration reading corresponding to 4 mA
Max. oxygen concentration C12 Oxygen concentration reading corresponding to 20 mA
Min. humidity C13 Humidity reading corresponding to 4 mA
Max. humidity C14 Humidity reading corresponding to 20 mA
Min. mixing ratio C15 Mixing ratio at 4 mA
Max. mixing ratio C16 Mixing ratio at 20 mA
Output damping coefficient C30 0 to 255 seconds
CAUTION
When you select logarithmic mode in Section 8.1.3, “Output Mode,” later in this
manual, the oxygen concentration, humidity reading, and mixing ratio remain constant at
0.1% O
2
, 0.1% H
2
O and 0.01 kg/kg respectively.