7.7.3 Minimum Current (4 mA) and Maximum Current (20 mA) Settings
This section describes how to set the humidity readings corresponding to 4 mA and 20
mA to 30% H
2
O and 80% H
2
O respectively.
Switch operation Display Description
ENT
A01 Display after the password has been entered.
ENT
C01 Set the humidity reading at 4 mA. Change the parameter
code to C13. Touch the [ ] key to switch to Group C.
ENT
C01 Touch the [ ] key to move the position of the digit that is
flashing to the right.
ENT
C11 Touch the [ ] key to enter the number 1.
ENT
C11 Touch the [ ] key to move the position of the digit that is
flashing to the right.
ENT
C13 Touch the [ ] key to enter the number 3.
ENT
000 Touch the [ENT] key to display the current set value. The
humidity 0% H
2
0 is now being displayed.
ENT
000 Touch the [ ] key to move the position of the digit that is
flashing to the right.
ENT
030 Touch the [ ] key to enter the number 3.
ENT
030 If you touch the [ENT] key, all the digits flash.
ENT
030 Touch the [ENT] key again to stop flashing.
ENT
C13 Touch the [ENT] key to return to the parameter code
selection display.
ENT
C13 Set the humidity reading at 20 mA. Touch the [ ] key to move
the position of the digit that is flashing to the right.
ENT
C14 Touch the [ ] key to change the number 3 in C13 to "4."
ENT
025 Touch the [ENT] key to display the current set value.
ENT
025 Touch the [ ] key to move the position of the digit that is
flashing to the right.
ENT
085 Touch the [ ] key to change the number 2 in C25 to "8."
ENT
085 Touch the [ ] key to move the position of the digit that is
flashing to the right.
ENT
080 Touch the [ ] key to change the number 5 in C85 to "0." The
number changes from 5 to 6 . to 9 to 0.
ENT
080 If you touch the [ENT] key, all the digits flash.
ENT
080 Touch the [ENT] key again to stop flashing.
ENT
C14 If you touch the [ENT] key again, the parameter code
selection display will appear.
ENT Basic
Touch the [ ] key together with the [ENT] key to return to the
panel basic panel display. (This is not required if you proceed to
display make another setting.)
(The displayed numeric characters indicate the humidity of
the measurement gas.)
T7701.EPS
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
The symbol ( ) indicates that the corresponding keys are being touched, and the light
characters indicate flashing.