Owner manual

IM 11M12A01-05E
10-15
10. Other Functions
Alarm-related Items in Group D
T10.6D.EPS
Code Item Tuning Engineering unit Default
setting
D01 Oxygen concentration, high-high 0 to 100 % O
2
100% O
2
alarm setpoint
D02 Oxygen concentration, high-limit 0 to 100 % O
2
100% O
2
alarm setpoint
D03 Oxygen concentration, low-limit 0 to 100 % O
2
0% O
2
alarm setpoint
D04 Oxygen concentration, low-low 0 to 100 % O
2
0% O
2
alarm setpoint
D05 Humidity, high-high alarm setpoint 0 to 100 %H
2
O 100% H
2
O
D06 Humidity, high-alarm setpoint 0 to 100 % H
2
O 100 % H
2
O
D07 Humidity, low-alarm setpoint 0 to 100 % H
2
O 0 % H
2
O
D08 Humidity, low-low alarm setpoint 0 to 100 % H
2
O 0 % H
2
O
D11 Mixing ratio, high-high alarm setpoint 0 to 1 kg/kg 1 kg/kg
D12 Mixing ratio, high-alarm setpoint 0 to 1 kg/kg 1 kg/kg
D13 Mixing ratio, low-alarm setpoint 0 to 1 kg/kg 0 kg/kg
D14 Mixing ratio, low-low alarm setpoint 0 to 1 kg/kg 0 kg/kg
D30 Oxygen concentration alarm 0 to 9.9 % O
2
0.1% O
2
hysteresis
D31 Humidity alarm hysteresis 0 to 9.9 % H
2
O 0.1 % H
2
O
D32 Mixing ratio alarm hysteresis 0 to 0.1 kg/kg 0.001 kg/kg
D33 Delayed alarm action 0 to 255 Seconds 3 seconds
D41 Oxygen concentration, 0: Not detected Not detected
high-high alarm detection 1: Detection
D42 Oxygen concentration, high-limit 0: Not detected Not detected
alarm detection 1: Detection
D43 Oxygen concentration, low-limit 0: Not detected Not detected
alarm detection 1: Detection
D44 Oxygen concentration, low-low 0: Not detected Not detected
alarm detection 1: Detection
D45 Humidity, high-high alarm detection 0: No detection No detection
1: Detection
D46 Humidity, high-alarm detection 0: No detection No detection
1: Detection
D47 Humidity, low-alarm detection 0: No detection No detection
1: Detection
D48 Humidity, low-low alarm detection 0: No detection No detection
1: Detection
D51 Mix ratio, high-high alarm 0: No detection No detection
1: Detection
D52 Mix ratio, high-alarm detection 0: No detection No detection
1: Detection
D53 Mix ratio, low-alarm 0: No detection No detection
1: Detection
D54 Mix ratio, low-low alarm detection 0: No detection No detection