Owner manual
IM 11M12A01-05E
10-14
Output-related Items in Group C
Code Item Tuning Engineering Default setting
unit
C01 Analog output 0: Oxygen concentration Humidity
1: Humidity
2: Mixed ratio
C03 Output mode 0: Linear Linear
1: Logarithm
C04 Output during warm-up 0: Held at 4 mA
up 1: Held at 20 mA
2: Set value remains held. Held at 4 mA.
C05 Output during 0: Not held
maintenance 1: Held output just before Held output just before
maintenance service. maintenance service.
2: Set value remains held.
C06 Output during 0: Not held Held output just before
calibration 1: Held output just before calibration.
calibration.
2: Set value remains held.
C07 Output in abnormal 0: Not held Held output at a preset
state 1: Held output just before value.
abnormal state occurs.
2: Set value remains held.
C11 Min. oxygen See Section 8.1. % O
2
0% O
2
concentration
C12 Max. oxygen See Section 8.1. % O
2
25% O
2
concentration
C13 Minimum humidity See Section 8.1. % H
2
O 0% H
2
O
C14 Maximum humidity See Section 8.1. % H
2
O 25% H
2
O
C15 Minimum mixing ratio See Section 8.1. kg/kg 0 kg/kg
C16 Maximum mixing ratio See Section 8.1. kg/kg 0.2kg/kg
C30 Output damping 0 to 255 Seconds 0 second
factor
C31 Set value during 2.4 to 21.6 mA 4 mA
warm-up
C32 Set value during 2.4 to 21.6 mA 4 mA
maintenance
C33 Set value during 2.4 to 21.6 mA 4 mA
calibration
C34 Set value in abnormal 2.4 to 21.6 mA 3.4 mA
state