Owner manual

IM 11M12A01-05E
10-11
10. Other Functions
Table 10.6 Resetting
Switch operation Display Brief description
. ` ENT
Err-01
If an error occurs, the error number and "------" are displayed
/------
alternately, as given on the left.
. ` ENT
PASSno
Hold down the [ENT] key for at least three seconds.
. ` ENT
0000
Touch the [ENT] key again to switch to the password entry display.
. ` ENT
1000
Enter the password 1102.
Intermediate switch operations omitted.
. ` ENT
1102
. ` ENT
A01
. ` ENT
G01
Change the parameter code to "G30."
. ` ENT
G01
. ` ENT
G30
. ` ENT All the digits Touch the [ENT] key to execute resetting.
light up.
T10.5aE.EPS
The symbol ( ) indicates that the key are light characters indicate "flashing".
"/" indicate that the characters are displayed alternately.