Owner manual

IM 11M12A01-05E
10-9
10. Other Functions
10.3 Initialization Procedure
Follow the table below to initialize parameters. The password for initialization is 1255.
Table 10.5 Initialization Procedure
Switch operation Display Description
>
`
ENT
F30
>
`
ENT
0000
>
`
ENT
1000
>
`
ENT
1000
>
`
ENT
1200
>
`
ENT
1200
>
`
ENT
1250
>
`
ENT
1250
>
`
ENT
1255
>
`
ENT
1255
>
`
ENT
USr Go
>
`
ENT
USr Go
>
`
ENT
F30
T10.5E.EPS
Enter the parameter code for the item to be initialized.
The following show an example of entering "F30."
(Previous needed operations are omitted.)
Touch the [ENT] key to switch to the password entry display.
Enter the password 1255 for initialization.
After you enter the password and then touch the [ENT] key, all the digits flash.
Touch the [ENT] key again to display "USR GO."
The symbol ( ) indicates that the key is being touched.
Light characters indicate that the digits are flashing.
Touch the [ENT] key once more.
All the digits then flash for two to three seconds, and data initialization starts.
The initialization is complete, and the parameter code selection display
then appears.
WARNING
Do not attempt to turn off the equipment power during initialization (while “USR GO”
is flashing).