Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<7. Startup>
7-7
IM 11M12A01-02E
7.5 Conrmation of Converter Type Setting
This converter can be used for both the Oxygen Analyzer and the Humidity Analyzer.
Before setting the operating data, be sure to check that the desired converter model has been
set.
Note that if the converter type setting is changed, the operating data that have been set are then
initialized and the default settings remain. To set the desired operating data, follow these steps:
(1) Press the [Setup] key.
(2) Use the [▼] key to select “Commissioning” and press the [Enter] key.
(3) In the password display, enter the [Enter] key. If the password is to be set again, enter the
new password (for details, see Section 8.6.5, Setting Passwords, later in this manual).
(4) The Commissioning display shown in Figure 7.6 appears. Select “Basic setup” and press
the [Enter] key.
(5) The Basic setup display shown in Figure 7.7 then appears. Conrm the currently set
converter type. If the Humidity Analyzer option /HS was selected at the time of purchase, the
converter was set for high temperature humidity use before shipment.
(6) If the converter type is to be changed, press the [Enter] key. The display shown in Figure 7.8
then appears.
(7) Use the [▼] key to select the type of equipment. Then press the [Enter] key to complete the
converter selection.
(8) If the type of converter is changed after setting the operating data, those data are then
initialized and the default settings remain. Reset the operating data to meet the new type of
equipment.
Enter
Commissioning
mA-output setup
Alarms setup
Contact setup
Others
Basic setup
F7.6E.ai
Enter
Basic setup
This instrument is for
Oxygen
Please note:changing this
Basic setting will return ALL
other setting to Default!!
Detector is ZR22
Select measure gas: Wet
F7.7E.ai
Figure 7.6 Commissioning Display Figure 7.7 Basic Setup
This instrument is for
Please note:changing this
Basic setting will return ALL
other setting to Default!!
Detector is ZR22
Select measure gas: Wet
Enter
Basic setup
Oxygen
Humidity
F7.8E.ai
Figure 7.8 Equipment Setup
8th Edition : Jan.13,2012-00