<5. Wiring>
5-12
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
5.8 Contact Input Wiring
The converter can execute specied function when receiving contact signals.
To use these contact signals, wire as follows:
ZR402G
Converter
Terminal box
DI-1
DI-C
DI-2
Contact input 1
Contact input 2
F5-14E.ai
Figure 5.13 Contact Input Wiring
5.8.1 Cable Specications
Use 2-core or 3-core cable for this wiring. Depending on the number of input(s), determine which
cable to use.
5.8.2 Wiring Procedure
(1) M4 screws are used for the terminals of the converter. Each cable should be terminated
corresponding to crimp-on terminals.
(2) The ON/OFF level of this contact input is identied by the resistance. Connect a contact
input that satises the specications in Table 5.2.
Table 5.2 Identication of Contact Input ON/OFF
Closed Open
Resistance 200 Ω or less 100 or more