Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<5. Wiring>
5-11
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
5.7 Wiring for ZR40H Automatic Calibration Unit
This wiring is for operating the solenoid valve for the zero gas and the span gas in the ZR40H
Automatic Calibration Unit, in a system where the calibration gas ow rate is automatically
controlled (e.g. System conguration 3). When installing this wiring, proceed as follows:
COMMON SPAN ZERO
to ZR402G terminal "AC-COM"
to ZR402G terminal "AC-SPAN"
to ZR402G terminal "AC-ZERO"
Wiring inlet
2-G1/2, Pg13.5, M20 or 1/2NPT
ZERO
SPAN
COMMON
Span gas solenoid valve
Zero gas solenoid valve
Terminal screw : M4
F5-12E.ai
Figure 5.11 Automatic Calibration Unit
5.7.1 Cable Specications
Use a 3-core cable for the above wiring.
5.7.2 Wiring Procedure
M4 screws are used for the terminals of the converter. Each cable should be terminated
corresponding to crimp-on terminals. M4 screws are used for the terminals of the solenoid valve
as well.
ZR402G
Converter
ZR40H Automatic
Calibration Unit
AC-Z
AC-C
AC-S
Zero
Span
F5-13.ai
Figure 5.12 Wiring for Automatic Calibration Unit