Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<5. Wiring>
5-10
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
5.5.2 Grounding Wiring
The ground wiring of the detector should be connected to the ground terminal of the detector
case. The ground wiring of the converter should be connected to either the ground terminal of the
converter case or the protective ground terminal in the equipment.
The grounding terminals of the detector and the converter are of size M4. Proceed as follows:
(1) Keep ground resistance to 100 Ω or less (equivalent JIS D style (Class 3)).
(2) When the ambient temperature of the wiring installation is 80 to 150 °C for the wiring of the
detector, use wiring material with sufcient heat resistance.
(3) When connecting the ground wiring to the ground terminal of the converter case, be sure
that the lock washer is in contact with the case surface (see Figure 5.9.).
(4) Ensure that the jumper plate is connected between the G terminal and the protective ground
terminal of the converter.
5.6 Contact Output Wiring
Contact outputs 1 to 3 can be freely assigned to “low limit alarm”, “high limit alarm”, etc. user
selectable, but the assignment of contact output 4 is xed (“error output”). And the action (contact
closed on error output) also cannot be changed.
When using these contact outputs, install the wiring as follows:
#1 Output
#2 Output
#3 Output
#4 Output
DO-1
DO-1
DO-2
DO-2
DO-3
DO-3
Terminal box Annunciator or the like
F5-11E.ai
ZR402G
Converter
DO-4
DO-4
Figure 5.10 Contact output wiring
5.6.1 Cable Specications
Number of wire in cable varies depending on the number of contact used.
5.6.2 Wiring Procedure
(1) M4 screws are used for the terminals of the converter. Each cable should be terminated
corresponding to crimp-on terminals.
(2) The capacities of the contact output relay are 30 V DC 3 A, 250 V AC 3 A. Connect a load
(e.g. pilot lamp and annunciator) within these limits.