Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<5. Wiring>
5-1
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
5. Wiring
In this Chapter, the wiring necessary for connection to the EXAxt ZR Separate type Zirconia
Oxygen Analyzer is described.
5.1 General
CAUTION
• NEVER supply current to the converter or any other device constituting a power circuit in
combination with the converter, until all wiring is completed.
• This product complies with CE marking.
Where compliance with CE marking is necessary, the following wiring procedure is necessary.
1.Install an external switch or circuit breaker to the power supply of the converter.
2. Use an external switch or circuit breaker rated 5 A and conforming with IEC 947-1 or
IEC 947-3.
3. It is recommended that the external switch or circuit breaker be mounted in the same room as
the converter.
4. The external switch or circuit breaker should be installed within the reach of the operator, and
marked as the power supply switch of this equipment.
Wiring procedure
Wiring should be performed according to the following procedure:
1. Be sure to connect the shield line to FG terminal of the converter.
2. The outer sheath of the signal line should be stripped to a length of 50 mm or less.
The most outer sheath of the power cable should be stripped to a length of 20 mm or less.
3. Signals may be affected by noise if signal lines, power cable and heater cable are located in
the same conduit. When using conduit, signal lines should be installed in a separate conduit
than power and heater cables.
4. Install metal blind plug(s) in unused cable connection gland(s) of the converter.
5. Metal conduit should be grounded.
6. The following cables are used for wiring:
Table 5.1 Cable specications
Terminal name of converter Name Need for shields Cable type Number of cores
CELL+, CELL-
HTR TC+, HTR TC-
CJ+, CJ-
Detector signal О CVVS 6
HEATER Detector heater CVV 2
L, N Power supply CVV 2 or 3 *
AO-1+, AO-1-, AO-2+, AO-2- Analog output О CVVS 2 or 4
DO-1, DO-2, DO-3, DO-4 Contact output CVV 2 to 8
AC-Z, AC-S, AC-C Automatic
Calibration unit
CVV 3
DI-1, DI-2, DI-C Contact input CVV 3
Note *: When the case is used for protective grounding, use a 2-core cable.