Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<4. Pipomg>
4-11
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
CAUTION
• As far as possible do not stop the instrument air ow, to prevent the sample gas from
entering the detector and damaging the zirconia cell.
• Connect the stop valve, which is at the calibration gas inlet, directly to the detector.
If there is piping between the detector and the valve, condensation may damage the sensor
by rapid cooling when calibration gas is introduced.
Figure 4.14 illustrates an example of System 2 using a detector with pressure compensation.
Supply air pressure (ow) may vary depending on the furnace pressure. It is recommended to
use a owmeter and an air set that is suitable for the furnace pressure.
NOTE
• When using the ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit, please note
that the supplying airow (pressure) will vary depending on the furnace pressure.
~
ZA8F Flow Setting Unit
EXA
ZR402G
ZR22G Separate type
Zirconia Oxygen Analyzer, Detector
ZR402G Converter
Reference
gas
Calibration
gas
Stop valve
or
Check valve
Needle
valve
Flowmeter
Instrument air
Air Set
Pressure
reducing
valve
Zero gas cylinder
Calibration gas
unit case
F4-13E.ai
Span gas
(Same as Zero gas calibration unit)
100 to
240 V AC
Contact input
Analog output, Contact output
Digital output (HART)
Signal
(6-core shield cable)
Heater (2 core cable)
Figure 4.14 System 2 using a detector with pressure compensation