Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<4. Pipomg>
4-8
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
4.3 Piping for System 3
Piping in System 3 is illustrated in Figure 4.10. In System 3, calibration is automated;
however, the piping is basically the same as that of System 2. Refer to Section 4.2.
Adjust secondary pressure of both the air set and the zero gas reducing valve so that these
two pressures are approximately the same. The ow rate of zero and span gases (normally
instrument air) are set by a single needle valve.
After installation and wiring, check the calibration contact output (see Sec. 7.11.2), and adjust
zero gas reducing valve and calibration gas needle valve so that zero gas ow is within the
permitted range. Next check span gas calibration contact output and adjust air set so that span
gas ow is within the permitted range.
CHECK
OUT
ZERO GAS IN
AIR IN
Instrument air Approx.
1.5 l/min.
Flowmeter
Flowmeter
REF
OUT
*2 *2
EV1 EV2
Solenoid valve
EV1, EV2
F2-6-3E.ai
*2: Needle valve is supplied as accessory with flowmeter
Figure 4.9 Piping inside in System 3
When blow back function is used by contact input to the ZR402G converter, install blow back
piping as illustrated in Section 4.3.1.
NOTE
Blow back function means the function to get rid of dust inside a probe in a high temperature
probe adapter by using compressed air, when a high temperature detector is used.
Instrument air
Pressure
reducing
valve
Air Set
Needle
valve
Flowmeter
Reference gas
Calibration gas
Check valve
~
Zero gas cylinder
ZR40H
Automatic Calibration Unit
ZR22G Separate type
Zirconia Oxygen Analyzer, Detector
ZR402G Converter
Calibration gas
unit case
F4-9E.ai
EXA
ZR402G
100 to 240 V AC
Contact input
Analog output, contact output
Digital output (HART)
Signal
(6-core shield cable)
Heater (2-core cable)
Figure 4.10 Piping for System 3