4-7
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
F4-8E.ai
Stop valve or Check valve
Piping for the calibration gas, 6 mm (O.D.)
by 4 mm (I.D.) Stainless steel pipe
Piping for the reference gas, 6 mm (O.D.)
by 4 mm (I.D.)Stainless steel pipe
Figure 4.8 Piping for the Calibration Gas Inlet
4.2.3 Piping for the Reference Gas
Reference gas piping is required between the air source (instrument air) and the ow setting unit,
and between the ow setting unit and the detector.
Insert the air set next to the ow setting unit in the piping between the air source and theow
setting unit.
Use a 6 mm (O.D.) x 4 mm or larger (I.D.) (or nominal size 1/4 inch) stainless steel pipe between
the ow setting unit and the detector.
4.2.4 Piping to the High Temperature Probe Adapter
Refer to Section 4.1.4 for the piping.