Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<4. Pipomg>
4-6
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
4.2.1 Piping Parts for System 2
Check that the parts listed in Table 4.2 are ready.
Table 4.2 Piping Parts
Detector Piping location Parts Name Note
General use
detector
Calibration gas inlet Stop valve or check valve Recommended by YOKOGAWA (L9852CB or
G7016XH) Provided by YOKOGAWA
(K9292DN or K9292DS)
Nipple * Rc1/4 or 1/4 NPT General parts
Zero gas cylinder User´s scope
Pressure reducing valve Recommended by YOKOGAWA (G7013XF or
G7014XF)
Joint for tube connection Rc1/4 or 1/4 NPT General parts
Reference gas inlet Air set Recommended by YOKOGAWA (G7003XF/
K9473XK or G7004XF/K9473XG)
Joint for tube connection Rc1/4 or 1/4 NPT General parts
High temperature
Detector
(0.15 m)
Calibration gas inlet Stop valve or check valve Recommended by YOKOGAWA (L9852CB or
G7016XH) Provided by YOKOGAWA
(K9292DN or K9292DS)
Nipple * Rc1/4 or 1/4 NPT General parts
Zero gas cylinder User´s scope
Pressure reducing valve Recommended by YOKOGAWA (G7013XF or
G7014XF)
Joint for tube connection Rc1/8 or 1/8 NPT
Reference gas inlet Air set Recommended by YOKOGAWA (G7003XF/
K9473XK or G7004XF/K9473XG)
Joint for tube connection Rc1/4 or 1/4 NPT General parts
Sample gas outlet Ejector assembly * Recommended by YOKOGAWA (E7046EC or
E7046EN)
T-shaped joint of the same
diameter *
R1/4 or 1/4 NPT General parts
Needle valve * Rc1/4 or 1/4 NPT General parts
Nipple of other diameter * R1/2 to R1/4 or R1/2 to 1/4
NPT
General parts
Note: Parts with marking * are used when required.
General parts can be found on the local market.
4.2.2 Piping for the Calibration Gas
This piping is to be installed between the zero gas cylinder and the ZA8F ow setting unit, and
between the ZA8F ow setting unit and the ZR22G detector.
The cylinder should be placed in a case assemble E7044KF or the like to avoid any direct
sunlight or radiant heat so that the gas cylinder temperature does not exceed 40 °C.
Mount a reducing valve (specied by YOKOGAWA) on the cylinder.
Mount a check valve or stop valve (specied by YOKOGAWA) on the nipple (found on the local
market) at the calibration gas inlet of the detector as illustrated in Figure 4.8.
(The check valve or the stop valve may have been mounted on the detector when shipped.)
Connect the ow setting unit and the detector to a stainless steel pipe 6 mm (O.D.) x 4 mm or
larger (I.D.) (or nominal size 1/4 inch).