<3. Installation>
3-10
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
3.4 Installation of ZA8F Flow Setting Unit
3.4.1 Installation Location
The following should be taken into consideration:
(1) Easy access to the unit for checking and maintenance work.
(2) Near to the detector or the converter
(3) No corrosive gas.
(4) An ambient temperature of not more than 55 °C and little changes of temperature.
(5) No vibration.
(6) Little exposure to rays of the sun or rain.
3.4.2 Mounting of ZA8F Flow Setting Unit
The ow setting unit can be mounted either on a pipe (nominal JIS 50A) or on a wall. It should be
positioned vertically so that the owmeter works correctly.
<Pipe Mounting>
(1) Prepare a vertical pipe of sufcient strength (nominal JIS 50A: O.D. 60.5 mm) for mounting
the ow setting unit. (The unit weighs approximately 2 to 3.5 kg.)
(2) Mount the ow setting unit on the pipe by tightening the nuts with the U-bolt so that the metal
tting is rmly attached to the pipe.
F3.12E.ai
Figure 3.12 Pipe Mounting
<Wall Mounting>
(1) Make a hole in the wall as illustrated in Figure 3.13.
(2) Mount the ow setting unit. Remove the pipe mounting parts from the mount ttings of the
ow setting unit and attach the unit securely on the wall with four screws.
223
140
F3-13E.ai
4 - Φ6.5 hole, or M6 screw
Unit : mm
F3-14E.ai
Figure 3.13 Mounting holes Figure 3.14 Wall mounting