Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<2. Specications>
2-25
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
2.7.6 Air Set
This set is used to lower the pressure when instrument air is used as the reference and span
gases.
Standard Specications
• G7003XF, K9473XK
Primary Pressure: Max. 1 MPa G
Secondary Pressure: 0.02 to 0.2 MPa G
Connection: Rc1/4 or 1/4FNPT with joint adapter
Weight: Approx.1 kg
Part No. Description
G7003XF Joint: Rc 1/4, Material: Zinc alloy
K9473XK Joint: 1/4 FNPT with adapter, Material: Zinc alloy, Adapter: SUS 316
• G7004XF, K9473XG
Primary Pressure: Max. 1 MPa G
Secondary Pressure: 0.02 to 0.5 MPa G
Connection: Rc1/4 or 1/4FNPT with joint adapter
Weight: Approx. 1 kg
Part No. Description
G7004XF Joint: Rc 1/4, Material: Zinc alloy
K9473XG Joint: 1/4 FNPT with adapter, Material: Zinc alloy, Adapter: SUS 316
l External Dimensions
Ø74
Primary
Secondary
88
Max. 210
max. 55
Approx. 122
Secondary
pressure
gauge
Panel (Horizontal mounting)
Panel (Vertical mounting)
G7003XF, G7004XF: Rc 1/4
K9473XK, K9473XG: 1/4NPT connector
airset.ai
A
22
ø15
2-ø2.2
+0.5
-0
2-ø6.5
40
2-ø6 screw depth 8
40
Horizontal
mounting
Vertical
mounting
Panel cut dimensions
View A
Unit : mm