Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<2. Specications>
2-23
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
Graph explanation
1) Graph 1 is to compensate for pressure loss in piping between the ejector and the pressure
gauge, and nd Po (pressure setting).
2) Graph 2 shows correlation between P (drive pressure) and Qa (air consumption).
3) Graph 3 shows correlation between P (drive pressure) and Pg (suction pressure; when the
sample gas inlet of the ejector is closed).
4) Graph 4 shows correlation between P (drive pressure) and Qg (suction ow) for each gas
pressure.
Qa
L
200
100
0
5
10
15
P= 0.5
0
40
30
40
60
80
Po (kPa)
Qa (l/min)
0
-1.0
-0.5
40
60
80
Pg (kPa)
0
8
4
40
60
80
Qg (l/min)
L (m)
P (kPa)
P (kPa)
P (kPa)
Po (kPa) : Pressure setting
P (kPa) : Drive pressure (at the ejector entrance)
Pg (kPa) : Suction pressure
Qa (l/min) : Air consumption
Qg (l/min) : Suction flow
L (m) : Distance between the ejector and the pressure
gauge
Needle
valve
Pressure gauge
Ejector
Po
Air
source
Gas
Pg
Qg
Gas pressure : -15 Pa
Air consumption characteristicsPressure setting characteristics
Suction pressure characteristics
Suction flow characteristics
F2-9E.ai
Gas Pressure:
0 kPa
1)
2)
3)
4)
Needle valve
<1>
<2>
Pressure gauge
Pressure gauge assembly
Nozzle (Note1)
R1/2
Ejector
Detector
<1> Rc1/4 or 1/4 FNPT
<2> Ø6/Ø4 or 1/4 inch copper tube (stainless)
with ejector to connect
<3> R1/4 or 1/4 NPT
40
20
Tee
Blow Rc1/4
Ø43
Approx. 67
Approx. 88 39
Approx. 70
38
Full open height
(Note1) The connecter of ejector assembly is a
dedicated connecter with nozzle function.
Instrument air inlet
<3>