Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<2. Specications>
2-21
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
2.7 Other Equipments
2.7.1 Dust Filter for the Detector (K9471UA)
This lter is used to protect the detector sensor from a corrosive dust components or high velocity
dust in recovery boilers and cement kilns.
Sample gas ow rate is needed to be 1 m/sec or more to replace gas inside zirconia sensor.
• Standard specication
Applicable detector: Standard-type detector for general use (the sample gas ow should be
approximately perpendicular to the probe.)
Mesh: 30 microns
Material: Carborundum (Filter), SUS316 (JIS)
Weight: Approx. 0.2 kg
Part No. Description
K9471UA Filter
K9471UX Tool
Ø51
32
10
Carborundum filter (SiC)
Detector
Screw
Increasing of insertion length
F3-2E.ai
Attach the filter unit to the tip of
the detector by screwing it clockwise.
2.7.2 Dust Guard Protector (K9471UC)
Recommended to be used when sample gas is likely to ow directly into the cell due to its ow
direction in the stack or the like, ammable dust may go into the cell, or water drops are likely to
fall and remain in the cell during downtime or the like due to the installation position.
Material: SUS316
Weight: Approx. 0.3 kg
4-Ø6
122
Ø50.8
100
Increasing of insertion length
F11-1E.ai