Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<2. Specications>
2-14
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
Zero calibration gas concentration setting range: 0.3 to 100 vol% O
2
(in increments of
0.01 vol% O
2
in smallest units).
Span calibration gas concentration setting range: 4.5 to 100 vol% O
2
(in increments of
0.01 vol% O
2
in smallest units).
Use nitrogen-balanced mixed gas containing 0 to 10 vol% O
2
scale of
oxygen, and 80 to 100 vol% O
2
scale of oxygen for standard zero gas
and standard span gas respectively.
Calibration interval; date/time setting: maximum 255 days
l Model and Codes
Model Sufx code
Option
code
Description
ZR402G - - - - - - - - - - - - - - - - - - - - - -
Separate type Zirconia Oxygen Analyzer, Converter
Converter
thread
-P
-G
-M
-T
- - - - - -
- - - - - -
- - - - - -
- - - - - -
G1/2
Pg13.5
M20x1.5 mm
1/2NPT
Display -J
-E
-G
-F
-C
- - - - - -
- - - - - -
- - - - - -
- - - - - -
- - - - - -
Japanese
English
German
French
Chinese
Instruction manual -J
-E
-C
- - - - - -
- - - - - -
- - - - - -
Japanese
English
Chinese
—
-A - - - - - - Always -A
Options
Tag plates
NAMUR NE43
compliant
/H Hood (*2)
/SCT
/PT
Stainless steel tag plate (*1)
Printed tag plate (*1)
/C2
/C3
Failure alarm down-scale:
Output status at CPU failure and hardware error is 3.6 mA or
less (*3)
Failure alarm up-scale:
Output status at CPU failure and hardware error is 21.0 mA or
more (*3)
*1 Specify either /SCT or /PT option code.
*2 Sun shield hood is still effective even if scratched.
*3 Output signal limits: 3.8 to 20.5 mA. Specify either /C2 or /C3 option code.
(Note) If AC line voltage is 125 V AC or greater, or in the EEC, the ZO21D cannot be used with the ZR402G.