Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<2. Specications>
2-10
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
F2-4E.ai
Rc1/2(Note2)
Approx. 351
180
Detector(ZR22G)
Ø52 over
Ø30
Ø60.5
Reference gas inlet Rc1/4 or 1/4NPT
Calibration gas inlet Rc1/4 or 1/4NPT
High temperature
Probe SiC pipe
110
48
25
170
Approx. 215
(Insertion length) (Note1)
Approx. 100
69
Approx. 48
C
(Note 1) L= 0.5、0.6、0.7、0.8、0.9、
1.0、1.5(m)
(Note 2) Sample gas outlet
(if the sample gas pressure is negative,
connect the auxiliary ejector.)
Ø124±3
85
Sample gas outlet
Flange provided
by customer
Unit: mm
L
t
Flange (Thickness 5)
JIS 5K 32 FF equivalent
Gasket (Thickness 1.5)
Flange <1>
Ø60.5
ØB
ØA
Pipe hole (2- G1/2, 2-1/2NPT, etc)
ØA
Ø115
Flange<1> A B C t
JIS 5K 50 FF SUS304 155 130 4 - Ø15 14
JIS 10K 65 FF SUS304 175 140 4 - Ø19 18
JIS 10K 80 FF SUS304 185 150 8 - Ø19 18
JIS 10K 100 FF SUS304 210 175 8 - Ø19 18
ANSI Class 150 4 RF SUS304 228.6 190.5 8 - Ø19 24
ANSI Class 150 3 RF SUS304 190.5 152.4 4 - Ø19 24
ANSI Class 150 2 RF SUS304 152.4 120.6 4 - Ø19 19
JPI Class 150 4 RF SUS304 229 190.5 8 - Ø19 24
JPI Class 150 3 RF SUS304 190 152.4 4 - Ø19 24
DIN PN10 DN50A SUS304 165 125 4 - Ø18 18