Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<2. Specications>
2-1
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
2. Specications
This chapter describes the specications for the following:
ZR22G General use separate type detector (See Section 2.2.1)
ZO21R Probe protector (See Section 2.2.2)
ZR22G (0.15 m) High temperature separate type detector (See Section 2.3.1)
ZO21P Adapter for High temperature probe (See Section 2.3.2)
ZR402G Separate type converter (See Section 2.4)
ZA8F Flow setting unit (See Section 2.5.1)
ZR40H Automatic calibration unit (See Section 2.5.2)
ZO21S Standard gas unit (See Section 2.6)
2.1 General Specications
2.1.1 Standard Specications
Measured Object: Oxygen concentration in combustion exhaust gas and mixed gas
(excluding inammable gases, may not be applicable corrosive gas such
as ammonia is present-check with YOKOGAWA)
The sampling gases containing a corrosive gas such as ammonia or
chlorine may be applicable to our oxygen gas analyzer. In this case,
contact with YOKOGAWA and its agency.
Measurement System: Zirconia system
Oxygen concentration: 0.01 to 100 vol% O
2
Output Signal: 4 to 20 mA DC (maximum load resistance 550 Ω)
Measurement Range: Any setting in the range of 0 to 5 through 0 to 100 vol% O
2
(in 1 vol% O
2
), or partial range
Digital Communication (HART): 250 to 550 Ω, depending on number of eld devices connected
to the loop (multi-drop mode).
Note: HART is a registered trademark of the HART Communication Foundation.
Display Range: 0 to 100 vol% O
2
Warm-up Time: Approx. 20 min.
Repeatability: (Excluding the case where the reference gas is by natural convection)
±0.5% Maximum value of set range. (range from 0 to 5 vol% O
2
or more
and less than 0 to 25 vol% O
2
range)
±1% Maximum value of set range.
(range from 0 to 25 vol% O
2
or more and up to 0 to 100 vol% O
2
range)
Linearity: (Excluding standard gas tolerance)
(Excluding the case where the reference gas is by natural convection)
(Use oxygen of known concentration (within the measuring range) as the
zero and span calibration gases.)
±1% Maximum value of set range ;
0 to 5 vol% O
2
or more and less than 0 to 25 vol% O
2
range
(Sample gas pressure: within ±4.9 kPa)
±3% Maximum value of set range ;
0 to 25 vol% O
2
or more and less than 0 to 50 vol% O
2
range