Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<12. Troubleshooting>
12-11
IM 11M12A01-02E
12.3 Countermeasures When Measured Value
Shows Error
The causes that the measured value shows an abnormal value is not always due to instrument
failures. There are rather many cases where the causes are those that measuring gas itself
is in abnormal state or external causes exist, which disturb the instrument operation. In this
section, causes of and measures against the cases where measured values show the following
phenomena will be described.
(1) The measured value is higher than the true value.
(2) The measured value is lower than the true value.
(3) The measured value sometimes shows abnormal values.
12.3.1 Measured Value Higher Than True Value
<Causes and Countermeasures>
(1) The measuring gas pressure becomes higher.
The measured oxygen concentration value X (vol% O
2
) is expressed as shown below,
when the measuring gas pressure is higher than that in calibration by ∆p (kPa).
X=Y [ 1+ (∆p/101.30) ]
where Y: Measured oxygen concentration value at the same pressure as in calibration
(vol% O
2
).
Where an increment of the measured value by pressure change cannot be neglected,
measures must be taken.
Investigate the following points to perform improvement available in each process.
• Is improvement in facility’s aspect available so that pressure change does not occur?
• Is performing calibration available under the average measuring gas pressure
(internal pressure of a furnace)?
(2) Moisture content in a reference gas changes (increases) greatly.
If air at the detector installation site is used for the reference gas, large change of moisture
in the air may cause an error in measured oxygen concentration value (vol% O
2
).
When this error is not ignored, use a gas in which moisture content is constant such as
instrument air in almost dry condition as a reference gas.
In addition, change of moisture content in exhaust gas after combustion is also considered
as a cause of error. However, normally this error is negligible.
(3) Calibration gas (span gas) is mixing into the detector due to leakage.
If the span gas is mixing into the detector due to leakage as a result of failure of the valve
provided in the calibration gas tubing system, the measured value shows a value a little
higher than normal.
Check valves (needle valves, check valves, solenoid valves for automatic calibration,
etc.) in the calibration gas tubing system for leakage. For manual valves, check them after
conrming that they are in fully closed states. In addition, check the tubing joints for leakage.
(4) The reference gas is mixing into the measuring gas and vice versa.
Since the difference between oxygen partial pressures on the sensor anode and cathode
sides becomes smaller, the measured value shows a higher value.
An error which does not appear as the Error1 may occur in the sensor. Sample gas and/or
the reference gas may be leaking. Visually inspect the sensor. If any crack is found, replace
the sensor assembly with a new one.
NOTE
Data such as cell robustness displayed in the Detailed-data display should also be used for
deciding sensor quality as references.
8th Edition : Jan.13,2012-00