Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<12. Troubleshooting>
12-10
IM 11M12A01-02E
12.2.2.6 Alarm 11: Thermocouple Voltage Alarm
This alarm is generated when the emf (voltage) of the thermocouple falls below -5 mV (about
-170 °C) or exceeds 42.1 mV (about 1020 °C). Whenever Alarm 11 is generated, Error2 (heater
temperature failure) occurs.
(1) Breakage of the heater thermocouple signal wire between the converter and the detector
occurred, or the cable is not securely connected to the connecting terminals.
(2) The positive and negative poles of the heater thermocouple signal wiring are shorted out in
the wiring extension or at the connection terminals.
(3) A failure of the thermocouple at the detector heater assembly occurred.
(4) A failure of the electrical circuits inside the converter occurred.
<Locating cause of failure, and countermeasures>
(1) Stop the power to the converter.
(2) Remove the wiring from terminals 3 and 4 of the detector and measure the resistance
between these terminals. If the resistance value is 5 Ω or less, the thermocouple seems to
be normal. If it is higher than 5 Ω, it may indicate the possibility that the thermocouple has
broken or is about to break. In this case, replace the heater unit (refer to Section 11.1.3,
Replacement of the Heater Unit).
CAUTION
• Measure the thermocouple resistance value after the difference between the detector tip
temperature and ambient temperature falls to 50 °C or less. If the thermocouple voltage is
large, accurate measurement cannot be achieved.
(3) If the thermocouple is normal, check whether or not the wiring cable is broken or shorted
out, and also whether the wiring cable is securely connected to the terminals.
Also check that the wiring resistance between the converter and the detector is 10 Ω or less.
(4) If there is no failure in the wiring, the electrical circuits inside the converter may possibly fail.
Contact the service personnel at Yokogawa.
12.2.2.7 Alarm 13: Battery Low Alarm
An internal battery is used as backup for the clock. After this alarm occurs, removing power from
the instrument may cause the clock to stop but should not affect stored parameters. The internal
clock is used for blow back scheduling; if you use this then after a battery alarm occurs (until the
battery is replaced) be sure to check / correct the date and time every time you turn on the power.
<Corrective action>
When the battery low alarm occurs, remember that the battery cannot be replaced by the user.
Contact your Yokogawa service representative.
NOTE
Battery life varies with environmental conditions.
* If power is applied to the instrument continuously, then the battery should not run down, and
life is typically about ten years. However the battery will be used during the time interval
between shipment from the factory and installation.
* If power is not applied to the instrument, at normal room temperatures of 20 to 25 °C then
battery life is typically 5 years, and outside this range but within the range -30 to +70 °C then
battery life is typically 1 year.
8th Edition : Jan.13,2012-00