Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<12. Troubleshooting>
12-5
IM 11M12A01-02E
12.2 Displays and Remedies When Alarms are
Generated
12.2.1 Alarm Types
When an alarm is generated, the alarm indication blinks in the display to notify of the alarm
(Figure 12.3). Pressing the alarm indication displays a description of the alarm.
Alarms include those shown in Table 12.2.
Tag:
21.0
%O
2
17.42mA -Output1
17.42mA -Output2
F12.3E.ai
Tag:
Alarm 6 :Zero conc.ratio
F12.4E.ai
Figure 12.3 Figure 12.4
Table 12.2 Types of Alarms and Reasons for Occurrence
Alarm
Code
Alarm Type Occurrence Conditions
Alarm 1 Oxygen concentration alarm Measured oxygen concentration value exceeds or falls below
the preset alarm limits. (refer to Section 8.3, Setting Oxygen
Concentration Alarms).
Alarm 6 Zero-point calibration coefcient
alarm
In automatic or semi-automatic calibration, zero correction
ratio is outside the range of 100 ±30%. (refer to Section 9.1.3,
Compensation).
Alarm 7 Span-point calibration coefcient
alarm
Span correction ratio is outside the range of 0±18% (refer to
Section 9.1.3, Compensation).
Alarm 8 EMF stabilization time-up alarm In automatic or semi-automatic calibration, cell voltage does not
stabilize after calibration time is up.
Alarm 10 Cold junction temperature alarm Temperature of the cold junction placed in the detector terminal
box falls below -25 ºC or exceeds 155 ºC.
Alarm 11 Thermocouple voltage alarm Generated when thermocouple voltage exceeds 42.1 mV (about
1020 ºC) or falls below -5 mV (about -170 ºC).
Alarm 13 Battery low alarm Internal battery needs replacement.
If an alarm is generated, actions such as turning off the heater power are not carried out. The
alarm is cancelled when the cause of the alarm is removed. However, Alarm 10 and/or Alarm 11
may be generated concurrently with Error2 (heater temperature error). In this case, the operation
when the error occurs has priority.
If the power to the converter is turned off after an alarm is generated and the converter is restarted
before the cause of the alarm has been removed, the alarm will be generated again. However,
Alarms 6, 7, and 8 (alarms related to calibration) are not generated unless calibration is executed.
8th Edition : Jan.13,2012-00