Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<10. Other Functions>
10-15
IM 11M12A01-02E
10.4 Reset
Resetting enables the equipment to restart. If the equipment is reset, the power is turned off
and then back on. In practical use, the power remains on, and the equipment is restarted under
program control. Resetting will be possible in the following conditions:
(1) Error 1 – if the cell voltage is defective
(2) Error 2 – if a temperature alarm occurs
(3) Error 3 – if the A/D converter is defective
(4) Error 4 – if an EEPROM write error occurs
For details on error occurrence, consult Chapter 12, Troubleshooting, later in this manual.
If any of the above problems occurs, the equipment turns off the power to the detector heater.
To cancel the error, reset the equipment following the steps below, or turn the power off and then
back on.
NOTE
Make sure that before resetting or restarting the power that there is no problem with the detector
or converter.
If a problem arises again after the resetting, turn the power off and troubleshoot the problem by
consulting the Troubleshooting chapter later in this manual.
To reset the equipment, follow these steps:
(1) Press the [Setup] key in the Basic panel display to display the Execution/Setup display.
(2) Choose “Reset”. The Reset display shown in Figure 10.13 appears.
(3) Choose “Start reset” and then press the [Enter] key to reset the equipment and the
equipment will then be in its warm-up state.
Enter
Start reset
Cancel reset
Reset
F10.13E.ai
Figure 10.13 Reset Display
8th Edition : Jan.13,2012-00