Owner manual

Table Of Contents
<10. Other Functions>
10-12
IM 11M12A01-02E
Table 10.5 Initialization Items and Default Values - Part 1
Item Initialization Parameter Default setting
Equipment
selection
Type of equipment Not initialized
Detector ZR22
Sample gas Wet gas
Displayed data
Display item
1st display item Oxygen concentration
2nd display item Current output 1
3rd display item Current output 2
Tag name Deleted
Trend graph
Parameter Oxygen concentration
Sampling interval 30 seconds
Upper limit (graph) 25 vol%O
2
Lower limit (graph) 0 vol%O
2
Automatic return time 0 min.
Language Not initialized
Calibration data Calibration setting
Mode Manual
Calibration procedure Span - zero
Zero gas concentration 1.00 vol%O
2
Span gas concentration 21.00 vol%O
2
Output hold time 10 min., 00 sec. (Semi_Auto)
Calibration time 10 min., 00 sec. (Semi_Auto)
Interval 30 days, 00 hr. (Auto)
Start date 01/01/00 (Auto)
Start time 0:00 (Auto)
Blow back Blow back setting
Mode No function (invalid)
(Output) hold time 10 min., 00 sec.
Blow back time 10 min., 00 sec.
Interval 30 days, 00 hr.
Start date 01/01/00
Start time 0:00
Current output
data
mA-output 1
mA-output 2
Parameter Oxygen concentration
Min. oxygen concentration 0 vol%O
2
Max. oxygen concentration 25 vol%O
2
Output damping constant 0 sec.
Output mode Linear
Output hold setting
Warm-up 4 mA
Set value of Warm-up 4 mA
Maintenance Previous value held
Set value of Maintenance 4 mA
Calibration, blow back Previous value held
Set value of Calibration, blow
back
4 mA
Error Preset value held
Set value of Error 3.4 mA
8th Edition : Jan.13,2012-00