Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<10. Other Functions>
10-12
IM 11M12A01-02E
Table 10.5 Initialization Items and Default Values - Part 1
Item Initialization Parameter Default setting
Equipment
selection
Type of equipment Not initialized
Detector ZR22
Sample gas Wet gas
Displayed data
Display item
1st display item Oxygen concentration
2nd display item Current output 1
3rd display item Current output 2
Tag name Deleted
Trend graph
Parameter Oxygen concentration
Sampling interval 30 seconds
Upper limit (graph) 25 vol%O
2
Lower limit (graph) 0 vol%O
2
Automatic return time 0 min.
Language Not initialized
Calibration data Calibration setting
Mode Manual
Calibration procedure Span - zero
Zero gas concentration 1.00 vol%O
2
Span gas concentration 21.00 vol%O
2
Output hold time 10 min., 00 sec. (Semi_Auto)
Calibration time 10 min., 00 sec. (Semi_Auto)
Interval 30 days, 00 hr. (Auto)
Start date 01/01/00 (Auto)
Start time 0:00 (Auto)
Blow back Blow back setting
Mode No function (invalid)
(Output) hold time 10 min., 00 sec.
Blow back time 10 min., 00 sec.
Interval 30 days, 00 hr.
Start date 01/01/00
Start time 0:00
Current output
data
mA-output 1
mA-output 2
Parameter Oxygen concentration
Min. oxygen concentration 0 vol%O
2
Max. oxygen concentration 25 vol%O
2
Output damping constant 0 sec.
Output mode Linear
Output hold setting
Warm-up 4 mA
Set value of Warm-up 4 mA
Maintenance Previous value held
Set value of Maintenance 4 mA
Calibration, blow back Previous value held
Set value of Calibration, blow
back
4 mA
Error Preset value held
Set value of Error 3.4 mA