Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<10. Other Functions>
10-6
IM 11M12A01-02E
0.0 vol%O
2
25.0 vol%O
2
10 min./div
12.3 vol%O
2
60 data
Preset upper limit
Preset lower limit
Time per division,
calculated by the set
sampling interval
Currently measured
value
Sampling period
F10.5E.ai
Figure 10.5 Plotting Graph for Sampling Interval
10.1.2.3 Setting Upper and Lower Limit Values on Graph
Set upper- and lower-limit values on the graph in the following procedure:
Press “Upper limit” in the Trend graph display. The numeric-data entry key appears.
Enter the upper-limit value. Also enter the lower-limit value in the same way. The allowable
settings for both upper-limit and lower-limit values range from 0 to 100 vol%O
2
.
10.1.2.4 Default Setting
When the analyzer is delivered, or if data are initialized, the set data are by default, as shown in
Table 10.3.
Table 10.3 Default Values for Graph Setting
Item Default Value
Parameter Oxygen concentration
Sampling interval 30 seconds
Upper limit 25 vol%O
2
Lower limit 0 vol%O
2
NOTE
If a rapid change in the measured value occurs during sampling, no sampled data are plotted on
the graph. Use the graph indication tentatively. Check the output current for accurate data.
8th Edition : Jan.13,2012-00