Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<9. Calibration>
9-8
IM 11M12A01-02E
NOTE
When setting calibration timing requirements, bear the following precautions in mind:
(1) If the calibration interval is shorter than the sum of hold (output stabilization) time plus
calibration time, the second calibration start time will conict with the rst calibration. In such
a case, the second calibration will not be conducted. (When both zero and span calibrations
are to be performed, the calibration time is double that required for a single (zero or span)
calibration.)
(2) For the same reason, if the calibration start time conicts with manual calibration or semi-
automatic calibration, the current calibration will not be conducted.
(3) If the calibration time conicts with maintenance service or blow back operations, calibration
will start after completing the maintenance service or blow back operations (see Section
8.2.1, earlier in this manual).
(4) If 000 days, 00 hours are set for the calibration intervals, only the rst calibration will be
conducted; a second or later calibration will not be conducted.
(5) If a past date is set to the calibration start day, no calibration will be conducted.
9.2.2 Default Values
When the analyzer is delivered, or if data are initialized, the calibration settings are by default, as
shown in Table 9.1.
Table 9.1 Default Settings for Calibration
Item Default Setting
Calibration mode Manual
Calibration procedure Span - zero
Zero gas (oxygen) concentration 1.00%
Span gas (oxygen) concentration 21.00%
Hold (output stabilization) time 10 minutes, 00 seconds
Calibration time 10 minutes, 00 seconds
Calibration interval 30 days, 00 hours
Start date 01 / 01 / 00
Start time 0:00
8th Edition : Jan.13,2012-00