Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<9. Calibration>
9-5
IM 11M12A01-02E
9.2 Calibration Procedures
NOTE
Calibration should be made under normal operating conditions (if the probe is connected to a
furnace, the analyzer will undergo calibration under the operating conditions of the furnace).
To make a precise calibration, conduct both zero and span calibrations.
9.2.1 Calibration Setting
The following sets forth the required calibration settings:
9.2.1.1 Mode
There are three calibration modes available:
(1) Manual calibration which allows zero and span calibrations or either one manually in turn;
(2) Semi-automatic calibration which lets calibration start with the touchpanel or a contact input,
and undergoes a series of calibration operations following preset calibration interval and
stabilization time.
(3) Automatic calibration which is carried out automatically following preset calibration interval.
Calibrations are limited by the following mode selection:
• When Manual calibration is selected:
Manual calibration only can be conducted. (This mode does not allow semi-automatic calibration
with a contact input nor automatic calibration even when its start-up time has reached.)
• When Semi-automatic calibration is selected:
This mode enables manual and semi-automatic calibrations to be conducted. (The mode,
however, does not allow automatic calibration even when its start-up time has reached.)
• When Automatic calibration is selected:
This calibration can be conducted in any mode.
To execute this calibration, follow these steps:
(1) Select the [Setup] key from the Basic panel display to display the Execution/Setup display.
Then select “Maintenance” from the Execution/Setup display.
(2) Select “Calibration setup” from the Maintenance display. Then select “Mode” from the
Calibration setup display (see Figure 9.4).
Now you can select “Manual”, “Semi_Auto”, or “Auto” calibration.