Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<8. Detailed Data Setting>
8-15
IM 11M12A01-02E
8.6 Other Settings
8.6.1 Setting the Date-and-Time
The following describe how to set the date-and-time. Automatic calibration or blow back works
following this setting.
Proceed as follows:
1) From the Basic panel display if you touch the [Setup] key, the Execution/Setup display
appears.
2) Select “Commissioning” and the Commissioning display appears.
3) Select “Others” and the display of Fig 8.12 appears.
4) Select “Clock” and the display of Fig 8.13 appears.
5) Select “Set date” and the numerical entry display appears. To enter June 21. 2004 , enter
210604. Touch the [Enter] key and you revert to the display of Fig 8.13.
6) Select “Set time”. Enter the time in 24-hour format. To set 2:30 pm, enter 1430 on the
numerical display and touch enter. Touch the [Enter] key. The clock starts from 00 seconds..
Clock
Averaging
Fuel setup
Purging
Unit setup
Passwords
Defaults
Enter
Others
F8.12E.ai
Enter
Set date: 1 3 / 0 7 / 0 0
Set time: 1 4 : 3 9
1 3 J ul. 2 0 0 0
1 4 : 3 9 : 5 5
Clock
F8.13E.ai
Figure 8.12 Other Settings Figure 8.13 Clock Display
8.6.2 Setting Periods over which Average Values Are
Calculated and Periods over which Maximum and
Minimum Values Are Monitored
The equipment enables the display of oxygen concentration average values and maximum and
minimum values under measurement (see Section 10.1.1, later in this manual).
The following section describes how to set the periods over which oxygen concentration average
values are calculated and maximum and minimum values are monitored.
8.6.2.1 Procedure
1) From the Basic panel display touch [Setup] key, and the Execution/Setup display appears.
2) Select “Commissioning” and the Commissioning display appears.
3) Select “Others” then Averaging. The display of Fig. 8.14 appears.
4) Select “Set period over which average is calculated” and numerical entry display appears.
To set 3 hours, enter 003. Entry range is 1 to 255 hours.
5) Select “Set period over which maximum and minimum is stored” and numerical entry display
appears. To set 48 hours, enter 048. Entry range is 1 to 255 hours.