Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<8. Detailed Data Setting>
8-14
IM 11M12A01-02E
8.5.2 Setting Procedure
This setting example shows how to set “When contact input 1 Opens, Start Semi_Auto
Calibration”.
1) From the Basic panel display touch [Setup] key, and the [Execution/Setup] display appears.
2) Select “Commissioning” and the Commissioning display appears.
3) Select “Contact Setup” then “Input contacts”.
4) Select “Input 1” . Function Selection window (Fig 8.11) appears.
5) Select “Calibration start”.
6) For Input contact 1 select (operate if) “Closed”. (You can select “Open” or “Closed”).
7) Change Closed to Open.
Input1: No function
Input1 contact is active at:
Closed
Input2: No function
Input2 contact is active at:
Closed
Enter
Input contacts
F8.10E.ai
Input1: No function
Input1 contact is active at:
Closed
Input2: No function
Input2 contact is active at:
Closed
Enter
Input contacts
No function
Cal. gas press. low
Range change
Calibration start
Process up set
Blow back start
F8.11.ai
Figure 8.10 Input Contacts Display Figure 8.11 Input Contacts Display
8.5.3 Default Values
All contact inputs are set to “No function”(disabled) prior factory shipment or after data
initialization..
8th Edition : Jan.13,2012-00