Owner manual

Table Of Contents
<8. Detailed Data Setting>
8-11
IM 11M12A01-02E
Table 8.5 Output Contact Settings
Item to be selected Brief description
Alarm
and
Error
settings
High-high limit alarm IfHigh-High alarm ON” is selected, contact output occurs when the
high-high limit is issued. To do this, it is required, in Alarms setup, that
the high-high alarm be set on beforehand (see Section 8.3).
High limit alarm If “High alarm ON” is selected, contact output occurs when the high
limit alarm is provided. To do this, it is required, in Alarms setup, that
the high limit alarm be set on beforehand (see Section 8.3).
Low limit alarm If “Low alarm ON” is selected, contact output occurs when the low
limit alarm is provided. To do this, it is required, in Alarms setup, that
the low limit alarm be set on beforehand (see Section 8.3).
Low-low limit alarm If “Low-Low alarm ON” is selected, contact output occurs when the
low-low limit alarm is issued. To do this, it is required, in Alarms setup,
that the low-low alarm be set on beforehand (see Section 8.3).
Calibration coefcient alarm If Calibration coefcient alarm is ON (enabled), then when a zero
calibration coefcient alarm (Alarm 6) or span calibration coefcient
alarm (Alarm 7) occurs then calibration coefcient alarm contact
output occurs (see Sec. 12.2.1)
Startup power stabilization
timeout alarm
If set ON then contact output occurs when startup power stabilization
timeout alarm (Alarm 8) occurs (see Sec.12.2.1)
Error If “Error ON” is selected, contact output occurs when an error is
issued. (See Chapter 12, Troubleshooting).
Other
settings
Warm-up If “Warm-up ON” is selected, contact output occurs during warm-up.
For the denition of Warm-up (see Section 8.2.1).
Output range change If “Range change ON” is selected, contact output occurs (“answer-
back signal to a range change signal”) while a range change signal is
applied to a contact input. To do this, it is required, in Input contacts
setup, that the range change be selected beforehand. For more on
this (see Section 8.5).
Calibration If “Calibration ON” is selected, contact output occurs during
calibration. For the denition of Under calibration (see Section 8.2.1).
Maintenance If “Maintenance ON” is selected, contact output occurs during
maintenance. For the denition of Under maintenance (see Section
8.2.1).
Blow back If “Blow back ON” is selected, contact output occurs during blow
back. For the denition of During blow back (see Section 8.2.1).
High limit temperature alarm Not supported by the oxygen analyzer.
Calibration gas press. low If “Cal. gas press. low ON” is selected, contact output occurs
(“answer-back signal to a calibration gas low pressure signal”)when
a calibration gas low pressure signal is applied to the contact input.
To do this, it is required, in Input contacts setup, that “Cal. gas press.
low” be selected beforehand. For more on this (see Section 8.5).
Process upset If “Process upset” is selected, contact output occurs (“answer-back
signal to a process upset signal) when the process upset signal is
applied to the contact input. To do this, it is required, in Input contacts
setup, that “Process upset” be selected beforehand (see Section 8.5).
Note: To provide an alarm with an output contact, be sure to make an alarm setting.
When using contact output as an answer-back signal for an input contact, be sure to make an input contact setting.
8th Edition : Jan.13,2012-00