Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<8. Detailed Data Setting>
8-11
IM 11M12A01-02E
Table 8.5 Output Contact Settings
Item to be selected Brief description
Alarm
and
Error
settings
High-high limit alarm If “High-High alarm ON” is selected, contact output occurs when the
high-high limit is issued. To do this, it is required, in Alarms setup, that
the high-high alarm be set on beforehand (see Section 8.3).
High limit alarm If “High alarm ON” is selected, contact output occurs when the high
limit alarm is provided. To do this, it is required, in Alarms setup, that
the high limit alarm be set on beforehand (see Section 8.3).
Low limit alarm If “Low alarm ON” is selected, contact output occurs when the low
limit alarm is provided. To do this, it is required, in Alarms setup, that
the low limit alarm be set on beforehand (see Section 8.3).
Low-low limit alarm If “Low-Low alarm ON” is selected, contact output occurs when the
low-low limit alarm is issued. To do this, it is required, in Alarms setup,
that the low-low alarm be set on beforehand (see Section 8.3).
Calibration coefcient alarm If Calibration coefcient alarm is ON (enabled), then when a zero
calibration coefcient alarm (Alarm 6) or span calibration coefcient
alarm (Alarm 7) occurs then calibration coefcient alarm contact
output occurs (see Sec. 12.2.1)
Startup power stabilization
timeout alarm
If set ON then contact output occurs when startup power stabilization
timeout alarm (Alarm 8) occurs (see Sec.12.2.1)
Error If “Error ON” is selected, contact output occurs when an error is
issued. (See Chapter 12, Troubleshooting).
Other
settings
Warm-up If “Warm-up ON” is selected, contact output occurs during warm-up.
For the denition of Warm-up (see Section 8.2.1).
Output range change If “Range change ON” is selected, contact output occurs (“answer-
back signal to a range change signal”) while a range change signal is
applied to a contact input. To do this, it is required, in Input contacts
setup, that the range change be selected beforehand. For more on
this (see Section 8.5).
Calibration If “Calibration ON” is selected, contact output occurs during
calibration. For the denition of Under calibration (see Section 8.2.1).
Maintenance If “Maintenance ON” is selected, contact output occurs during
maintenance. For the denition of Under maintenance (see Section
8.2.1).
Blow back If “Blow back ON” is selected, contact output occurs during blow
back. For the denition of During blow back (see Section 8.2.1).
High limit temperature alarm Not supported by the oxygen analyzer.
Calibration gas press. low If “Cal. gas press. low ON” is selected, contact output occurs
(“answer-back signal to a calibration gas low pressure signal”)when
a calibration gas low pressure signal is applied to the contact input.
To do this, it is required, in Input contacts setup, that “Cal. gas press.
low” be selected beforehand. For more on this (see Section 8.5).
Process upset If “Process upset” is selected, contact output occurs (“answer-back
signal to a process upset signal) when the process upset signal is
applied to the contact input. To do this, it is required, in Input contacts
setup, that “Process upset” be selected beforehand (see Section 8.5).
Note: To provide an alarm with an output contact, be sure to make an alarm setting.
When using contact output as an answer-back signal for an input contact, be sure to make an input contact setting.
8th Edition : Jan.13,2012-00