Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<8. Detailed Data Setting>
8-8
IM 11M12A01-02E
Parameter: Oxygen
Hysteresis: 0.1%O
2
Contact delay: 3 s
Setpoints
Enter
Alarms setup
F8.4E.ai
High High alarm: OFF
Set value: 1 0 0 . 0 % O
2
High alarm: ON
Set value: 1 0 0 . 0 % O
2
Low alarm: OFF
Set value: 0 . 0 % O
2
Low Low alarm: OFF
Set value: 0 . 0 % O
2
Enter
Oxygen alarms
F8.5E.ai
Figure 8.4 Alarms Setup Display Figure 8.5 Oxygen Alarms Display
8.3.4 Default Values
When the analyzer is delivered, or if data are initialized, the default alarm set values are as
shown in Table 8.3.
Table 8.3 Alarm Setting Default Values
Item Setting range Default setting
Hysteresis 0 to 9.9 vol%O
2
0.1 vol%O
2
Delay time 0 to 255 seconds 3 seconds
High-high limit alarm OFF
High-high limit alarm setpoint 0 to 100 vol%O
2
100 vol%O
2
High limit alarm OFF
High limit alarm setpoints 0 to 100 vol%O
2
100 vol%O
2
Low limit alarm OFF
Low limit alarm setpoint 0 to 100 vol%O
2
0 vol%O
2
Low-low limit alarm OFF
Low-low limit alarm setpoint 0 to 100 vol%O
2
0 vol%O
2
8th Edition : Jan.13,2012-00