Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<8. Detailed Data Setting>
8-6
IM 11M12A01-02E
8.3 Oxygen Concentration Alarms Setting
The analyzer enables the setting of four alarms — high-high, high, low, and low-low alarms —
depending upon the oxygen concentration. The following section sets out the alarm operations
and setting procedures.
8.3.1 Setting the Alarm Values
(1) High-high and high alarm values
If high-high and high alarm values are set to ON, then alarms occur if measured valves exceed
the alarm set values. The oxygen alarm set values can be set in the range of 0 to 100 vol%O
2
.
(2) Low and low-low alarm values
If low-low and low alarm values are set, then alarms occur if measured values fall below the
alarm set values. The oxygen alarm set values can be set in the range of 0 to 100 vol%O
2
.
8.3.2 Alarm Output Actions
If the measured values of the oxygen concentration uctuate between normal (steady state)
values and the alarm setting, alarm outputs may be frequently issued and canceled.
To avoid this, set the alarm delay and hysteresis for alarm canceling under the alarm output
conditions, as Figure 8.3 shows.
When a delay time is set, an alarm will not be issued so quickly even if the measured value differs
from the steady state and enters the alarm setpoint range.
If the measured value remains within the alarm setpoint range for a certain period of time (for the
preset delay time), an alarm will be issued. On the other hand, there will be a similar delay each
time the measured value returns to the steady state from the alarm setpoint range (canceling the
alarm status).
If hysteresis is set, alarms will be canceled when the measured value is less than or greater than
the preset hysteresis values.
If both the delay time and hysteresis are set, an alarm will be issued if the measured value is in
the alarm setpoint range and the delay time has elapsed.
For the alarm to be reset (canceled), the measured value must be beyond the preset hysteresis
value and the preset delay time must have elapsed.
Refer to Figure 8.3 for any further alarm output actions. The delayed time and hysteresis settings
are common to all alarm points.
Delay time:
5 seconds
7.5%
High limit alarm setpoint
5.5%
Oxygen concentration
Alarm output
ON
OFF
Hysteresis
2.0%
A
B
C
D
Alarm range
Delay time:
5 seconds
Delay time:
5 seconds
F8.3E.ai
Figure 8.3 Alarm Output Action
8th Edition : Jan.13,2012-00