Owner manual

Table Of Contents
<8. Detailed Data Setting>
8-5
IM 11M12A01-02E
8.2.4 Default Values
When the analyzer is delivered, or if data are initialized, output hold is the default as shown in
Table 8.2.
Table 8.2 Output Hold Default Values
Status Output hold setting Preset value
During warm-up 4 mA 4 mA
Under maintenance Holds output at value just before maintenance started. 4 mA
Under calibration or blow back Holds output at value just before starting calibration or
blow back.
4 mA
On Error occurrence Holds output at a preset value. 3.4 mA
8th Edition : Jan.13,2012-00