Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<8. Detailed Data Setting>
8-2
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
Setting example 2
If the range minimum (corresponding to 4 mA output) is set to 75 vol%O
2
then range maximum
(corresponding to 20 mA output) must be at least 75x1.3=98 vol%O
2
(rounding decimal part up).
5
15
25
35
45
55
65
75
85
95
5 15 25 35 45 55 65 75
Minimum oxygen concentration, vol%O
2
(for a minimum current of 4 mA)
Minimum-Maximum setting range of oxygen concentration
Maximum oxygen concentration, vol%O
2
(for a maximum current of 20 mA)
Ranges over which oxygen
concentrations can be set
Outside ranges
F8.0E.ai
8.1.3 Setting Output Smoothing Coefcient
When the oxygen concentration in the sample gas uctuates rapidly, if the measured value is
used for control this can lead to problems with undesirable frequent ON/OFF switching. You can
set a smoothing time constant of between 0 and 255 seconds to reduce the effect. Select the
appropriate output damping constant from the numeric-data entry display. To set 30 seconds,
enter 030.
8.1.4 Selection of Output Mode
You can select whether the relationship between the sample oxygen concentration and the
analog output signal be linear or logarithmic. Press the [Enter] key in the output mode display. A
linear/ logarithmic selection display then appears. Select the desired mode.
NOTE
If you select an output mode of “logarithmic” then regardless of range setting the minimum output
value becomes 0.1 vol%O
2
xed.
8.1.5 Default Values
When the analyzer is delivered or reset to defaults, the output current default settings by as
shown in Table 8.1.1.
Table 8.1.1 Output Current Default Values
Item Default setting
Min. oxygen concentration 0 vol%O
2
Max. oxygen concentration 25 vol%O
2
Output damping constant 0 (seconds)
Output mode Linear