Owner manual
Table Of Contents
- Introduction
- CONTENTS
- 1. Overview
- 2. Specifications
- 2.1 General Specifications
- 2.2 General use Separate type Detector and Related Equipment
- 2.3 Separate type Detector for High Temperature and Related Equipment
- 2.4 ZR402G Separate type Converter
- 2.5 ZA8F Flow Setting Unit and ZR40H Automatic Calibration Unit
- 2.6 ZO21S Standard Gas Unit
- 2.7 Other Equipments
- 2.7.1 Dust Filter for the Detector (K9471UA)
- 2.7.2 Dust Guard Protector (K9471UC)
- 2.7.3 Ejector Assembly for High Temperature (E7046EC, E7046EN)
- 2.7.4 Stop Valve (L9852CB, G7016XH)
- 2.7.5 Check Valve (K9292DN, K9292DS)
- 2.7.6 Air Set
- 2.7.7 Zero Gas Cylinder (G7001ZC)
- 2.7.8 Cylinder Pressure Reducing Valve (G7013XF, G7014XF)
- 2.7.9 Case Assembly for Calibration Gas Cylinder (E7044KF)
- 2.7.10 ZR22A Heater Assembly
- 3. Installation
- 3.1 Installation of General-purpose Detector
- 3.2 Installation of High Temperature Detector (ZR22G-015)
- 3.3 Installation of the ZR402G Converter
- 3.4 Installation of ZA8F Flow Setting Unit
- 3.5 Installation of ZR40H Automatic Calibration Unit
- 3.6 Installation of the Case Assembly(E7044KF)
- 3.7 Insulation Resistance Test
- 4. Piping
- 5. Wiring
- 6. Components
- 7. Startup
- 7.1 Checking Piping and Wiring Connections
- 7.2 Checking Valve Setup
- 7.3 Supplying Power to the Converter
- 7.4 Touchpanel Switch Operations
- 7.5 Confirmation of Converter Type Setting
- 7.6 Confirmation of Detector Type Setting
- 7.7 Selection of Sample Gas
- 7.8 Output Range Setting
- 7.9 Setting Display Item
- 7.10 Checking Current Loop
- 7.11 Checking Contact I/O
- 7.12 Calibration
- 8. Detailed Data Setting
- 9. Calibration
- 10. Other Functions
- 11. Inspection and Maintenance
- 12. Troubleshooting
- Customer Maintenance Parts List CMPL 11M12A01-02E
- Customer Maintenance Parts List CMPL 11M12C01-01E
- Customer Maintenance Parts List CMPL 11M12A01-11E
- Customer Maintenance Parts List CMPL 11M03B01-10E
- Customer Maintenance Parts List CMPL 11M03B01-05E
- Customer Maintenance Parts List CMPL 11M03D01-01E
- Revision Information

<8. Detailed Data Setting>
8-1
IM 11M12A01-02E 8th Edition : Jan.13,2012-00
8. Detailed Data Setting
8.1 Current Output Setting
This section describes setting of the analog output range.
8.1.1 Setting Minimum Current (4 mA) and Maximum
Current (20 mA)
To set the minimum and maximum currents, proceed as follows:
(1) Select [Commissioning] in the Execution/Setup display.
(2) Select the “mA-output setup” in the Commissioning display.
(3) Select “mA-output1” in the mA-outputs display.
(4) Select “Min. oxygen conc” in the mA-output1 range display and press the [Enter] key. The
numeric-data entry display then appears. Enter the oxygen concentration for the minimum
current (4 mA); for example, enter “010” for 10 vol%O
2
.
(5) Select “Max. oxygen conc” in the mA-output1 range display and enter the oxygen
concentration for the maximum current (20 mA) in the same manner as in step 4 above.
(6) Set mA-output2 in the same way as the setting procedure for mA-output1 given above.
8.1.2 Input Ranges
The range low and high values are restricted as follows:
The range min. O
2
concentration value (corresponding to 4 mA output) can be set to either 0
vol%O
2
or in the range of 6 to 76 vol%O
2
.
The range max. O
2
concentration value (corresponding to 20 mA output) can be set to any value
in the range of 5 to 100 vol%O
2
, however the range max. setting must be at least 1.3 times the
range min. setting.
If you do not observe this restriction, the measurement will be invalid, and any previous valid
value will be used. The gray area in gure represents the valid setting range.
Setting example 1
If the range minimum (corresponding to 4 mA output) is set to 10 vol%O
2
then range maximum
(corresponding to 20 mA output) must be at least 13 vol%O
2
.