User guide

7-2
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<7. ROUTINE MAINTENANCE>
Figure 57 - Window Cleaning
TDLS200 TDL Analyzer Instruction Manual V2.1
Page 88 of 131
- 88 -
Window surface to
be cleaned
If the windows are contaminated they may be cleaned using the following procedure:
Remove analyzer from process (isolate, etc. if necessary)
Use a clean, dry instrument air or nitrogen pressure supply to first blow off any particulate
matter.
Using warm water and mild soap detergent, gently clean the window surface with a soft,
non-abrasive cloth.
If the deposits do not come off then use a small amount of IPA (Isopropyl Alcohol) and a
soft, non-abrasive cloth.
Use the same clean, dry instrument air or nitrogen pressure supply to blow dry the surface.
Carefully check the entire surface of the window from different angles to ensure it is
thoroughly cleaned and ready for service.
If the window does not appear to clean up well, then replace the window assembly with a
new one.
If the window appears to have an etched surface then it has probably been contaminated
with HF or other similar corrosive gas.
Figure 32 - Window Cleaning
Figure 56 - Window Replacement
TDLS200 TDL Analyzer Instruction Manual V2.1
Page 89 of 131
- 89 -
Figure 33 - Window Replacement
5.1.2 Replacing Windows:
If it has been determined that a window required replacement, then the following should be
performed:
Shut down the analyzer and remove power, thus ensuring the laser if OFF.
Remove Launch or Detect (as appropriate) from the process interface by removing the one
fixed hole fastener (5/32” Allen Wrench), loosening the remaining three fasteners and then
twisting and pull off the unit.
Remove the four retaining screws (3/32” Allen Wrench, T-Handle) and lock washers from
the window holder recessed inside the quick connect body.
Remove the window that requires replacement.
Ensure the new replacement window is of the same specification as the window being
removed. Standard Windows are 0.25” Thick wedged, BK-7 borosilicate crown glass
mounted in stainless steel retaining ring.
The window holder assemblies can only be installed in one way by virtue of the mounting
hole placement.
Ensure a new O-Ring (of same specification) is installed at the same time. Viton is the
standard material however, certain applications may require other materials – check original
specifications.
With the new O-Ring in position, carefully locate the new window holder in place with the
smooth glass surface against the O-Ring.
Carefully tighten all four screws (ensuring the lock washers are in place) using a T-handle
wrench. Keep tightening all the fasteners in turn by ¼ turns thus ensuring an even torque
loading.
Hand tighten the screws until they are very firmly tightened (a T handle wrench as shown
below will torque its own shaft by about ¼ turn when the screws are fastened securely).
Offset screw position
f
o
r win
do
w
O-ring correctly
seated
in
pos
i
t
i
o
n
O- ring correctly
seated in position