User guide

6-1 6-2
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<6. BASIC OPERATION>
Online Menu Level 1 Menu Level 2 Menu Level 3 Menu Level 4 Menu Level 5 Menu
Advanced
* Password
Protected
Congure
System I/O Analog Output Channel 1
Channel 2*
(Always uses track
mode)
Channel 3*
*(similar to Channel 1)
Warning Mode
Fault Mode*
*(similar to Warning)
Block Mode
Field Loop Check
AO CH Calibration
Conc1/Conc2/Tran/Temp/Pres/
None
4 mA- 20 mA
Block Mode (mA value)
Track mode
Hold Mode (mA value)
High (20 mA)
Low (3.3 mA)
CH 1 check, mA value
CH 2 check, mA value
CH 3 check, mA value
CH 1 Calibration
CH 2 Calibration
CH 3 Calibration
System I/O
Digital Output Warnings Detector signal low
Transmission Low
Spectrum noise hig
Process pressure out of range
Process temperature out of range
Concentration out of range
Board temperature out of range
DO-Warning delay
Faults Laser temperature out of range
Detector signal high
Dectecor signal lost
Measurement peak no response
Peak center out of range + 2nd gas
threshold
Walidation failure + base concen-
trations
DO-fault delay
User Alarm Conc/ Trans/ Val/ Cal
High/Low and limit
DO-use alarm delay
System Serial Number
Laser Serial Number
Password Old Password
New password Conrmation of Change
Software Version
Date & Time New Date
New Time
System Temperature Launch Unit (˚C)
Detect Unit (˚C)
TCP/IP
Set new IP address, sub-
net mask and
default gateway
Adjustable Resistors Laser
Detect - R21
Detect - R22
Detect - R23
Valve Control Valve 1
Valve 2*
Valve 3*
*(similar to valve 1)
Manual On/Off
Time Sequence Next Valve Selection
Valve-on duration in minutes
Restore Control Remote control channel
Signal
Processing
Laser Spectra & Control Gas 1 Concentration
Gas 2 Concentration
Gas temperature
Gas pressure
Transmission
Laser Temperature in ˚C
Peak Position