User guide

TDLS200 TDL Analyzer Instruction Manual V2.1
Page 19 of 131
- 19 -
2.3 Analyzer Components
Figure 1 - System Overview
Launch to Detect Interconnect (cable)
Launch Unit:
Main Electronics Housing
User Interface (optional)
Laser Assembly
Check Gas Flow Cell (for
On-Line Validation)
Detect Unit:
Detect Electronics
Housing
Detector
Assembly
Process Interface:
Analyzer detachable from process interface for Off-Line
calibration / service.
Flanged O-Ring Alignment
Flanged Metal Bellows Seal Alignment
Flow Cell
Isolation Flanges
By-Pass Piping
Custom designs for specific applications.
Hazardous Area Purge (optional)
NEC/CSA Class 1, Div. 2, Gr. A-D
ATEX Zone 2 Cat 3
Figure 4 - System Overview
4-1
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<4. ANALYZER COMPONENTS>
Launch Unit:
• Main Electronics Housing
• User Interface (optional)
• Laser Assembly
• Check Gas Flow Cell (for
On-Line Validation)
Hazardous Area Purge (optional)
• NEC/CSA Class 1, Div. 2, Gr. A-D
• ATEX Zone 2 Cat 3
Launch to Detect Interconnect (cable)
Launch Unit:
• Analyzer detachable from process
interface for Off-Line calibration / service.
• Flanged O-Ring Alignment
• Flanged Metal Bellows Seal Alignment
• Flow Cell
• Isolation Flanges
• By-Pass Piping
• Custom designs for specic applications.
Detect Unit:
• Detect Electronics
Housing
• Detector Assembly
• The Launch Unit and Detect Unit are connected to each other via a Tray Rated 4-pair shielded twisted
pair cable.
• The Launch Unit requires a single 24VDC power supply (by customer or via optional Power Supply
Unit).
• Nitrogen purge gas is required to prevent ambient oxygen ingress however, for other target gases it
may be possible to use Instrument Air for purging.
• The Process Interfaces are available in various formats, sizes and materials to suit the desired
measurement/installation.
• The available Remote Interface Unit (RIU) can be located typically up to 100m (330ft) away from the
Launch Unit. The RIU also requires a 24VDC power supply. The RIU connects to the Launch Unit on
Ethernet (10-base-T 10/100) via CAT5e eld rated cable.
• The available Universal Remote Display (URD) can be located typically up to 40m (120ft) away from the
Launch Unit. The URD requires an AC power input that is connected to a universal power supply with
24VDC output power supply (for the analyzer). The URD connects to the Launch Unit via multipair
sheilded twisted pair cable.
4 ANALYZER COMPONENTS