User guide

Performance Specification
Repeatability: Application Dependent
Linearity: +/- 1% of FS
Response time: 2-20 seconds, plus transport time for
extractive systems when applicable
Drift: Application Dependant
Installation Specifications
Hazardous Area: Zone 1: Contact Yokogawa
Zone 2: ATEX group II Cat. 3G with purge
system EEx pz II T5 (-20< Ta <50C)
By Design: Non-Hazardous Area; Purge required for
ATEX zone 1&2 and NEC Class 1
Division 1&2
Maximum Distance between Launch and Detect:
30 m (±90ft)
Maximum interconnecting cable 50m
Wetted Parts: Analyzer & standard Alignment Flange - 316
SS, BK-7 Glass, Teflon encapsulated Viton
and Silicone RTV sealant.
Optional: Isolation Flanges and Flow Cells - 316 SS,
Sapphire, Kalrez -
Also available in Monel A400, Hastelloy
C-276, Carpenter 20, Titanium Grade 2 and
others on request.
Utilities:
Instrument Air may be used as a purge gas in principle for all of the
below applications, but this will depend on the application type and the
required precision of the measurement.
Oxygen Analyzer N
2
CO Analyzer N
2
or Instrument Air
CO
2
Analyzer N
2
or other non-CO
2
containing inert gas
H
2
O ppm Analyzer N
2
with <20ppm levels H
2
O for feed to
optional Dryer Package
H
2
O % Analyzer N
2
Flow Rate: 5-30 L/min for window purge
2 L/min for validation, calibration and optical
purge
SIL Assessment:
The TDLS200 has a FMEDA assessment by exida and is
classified as a Type B1 device in compliance with the following
standards; IEC 61508 or EN 954-1. Functional Safety of Electrical/
electronic/programmable electronic related systems; SIL 1
capability for single device.
* The TDLS200 is not SIL certified as standard; to be certified the unit
must be specified and designed from the beginning to meet all SIL
specifications.
Basic System Configuration
The TruePeak can be installed in a number of ways depending on
process requirements. The most typical installation types are shown
below, however other installation methods are possible, please contact
Yokogawa with your application details.
Cross Stack/Pipe Configuration
Measures directly across process pipe or vessel
Typically has nitrogen or other purge gas protecting
process windows
Span Validation via serial flow cell (see Operation Specifications).
• Full calibration requires removal from process
May require pressure and temperature inputs
(Application Dependant)
Multiple methods to increase Optical Path Length (OPL) if needed
• 5 meter interconnection cable standard
Close Coupled Extractive / Bypass Configuration*
Measures across a section of pipe where process flow
is directed
The measurement section can be isolated from process flow for full
calibration/validation, zero and span
Process pressure and temperature can be controlled
or the analyzer may require pressure and temperature inputs
(Application Dependant)
Length of measurement section dependant on accuracy requirements
and process conditions
Extractive Configuration*
Sample is fully extracted from process
(and may be conditioned before measurement).
Flow cells are available with ability to purge in front of windows
(balanced flow cell) if required.
Process pressure and temperature can be controlled
or the analyzer may require pressure and temperature inputs
(Application Dependant)
Length of flow cell dependant on accuracy requirements
and process conditions
* Contact Yokogawa for further details
Zero check gas,
at grade
Span check gas.
at grade
Flow-Cell-Outlet
process
sample
inlet
Cal
inlet
PROCESS
3-1 3-2
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<3. GENERAL SPECIFICATIONS>