<10. Data Files anD Format>
analyzer to capture spectrum les and under what
condition. The default condition is related to the number
of measurement however, the user can select Relative or
Absolute changes pending the site specic conditions/
requirements.
Typical updates between capture value would be 300
The more frequently spectrum are stored then the larger
the MMDDYY.spe les will become.
NOTE: Capturing every spectrum for one day can create a
single day le in excess of 30MB – Choose the parameters
carefully to prevent memory overload!
Determine whether or not the analyzer should capture
spectrum les under a WARNING condition.
Note, this may be useful to do so however, if the Warning
alarm conditions are not set correctly then there could be
excessive les created for less meaningful Warning alarm
conditions.
An example is a low transmission warning alarm set at
70% for an application that often runs at less than 70%
transmission.
Determine whether or not the analyzer should capture
spectrum les under a FAULT condition.
Note, this may be useful to do so however, if the Fault
alarm conditions are not set correctly, then there could
be excessive les created for less meaningful Fault alarm
conditions.
These les are often useful for Factory based diagnostics.