<10. Data Files anD Format>
alarms .his Alarms History
A historical log of analyzer
alarm
events
Alarm History - in the form of ASCII data les that can
be opened with Microsoft “Notepad” as simple .txt le
formats. The content can then be copied and pasted into
Microsoft Excel spread. Each Alarm incident is logged
with the date (MM-DDYYYY) and time that it occurred, the
Mode
(Warning or Fault or User Alarm), the specic condition
within the Mode (e.g. Detector Signal Lost) and the state
(ON or OFF).
Example:
10-10-2007 13:21:36 Fault(peak center out of range) OFF
10-10-2007 19:00:25 Warning(detector signal low) ON
alarms .bak
Alarms History Back-Up
A back-up of the historical
log of
analyzer alarm events
larm History Back-up - When the alarms.his le exceeds
100KB size the contents is saved to this .bak le and the
.his le is emptied. This .bak le is in the form of ASCII
data les that can be opened with Microsoft “Notepad” as
simple .txt le formats. The content can then be copied
and pasted into Microsoft Excel spread. Each Alarm inci-
dent is logged with the date (MM-DD-YYYY) and time that
it occurred, the Mode (Warning or Fault or User Alarm), the
specic condition within
the Mode (e.g. Detector Signal Lost) and the state (ON or
OFF). Example:
10-10-2007 13:21:36 Fault(peak center out of range) OFF
10-10-2007 19:00:25 Warning(detector signal low) ON
calibr .his
Calibration History
A historical log of analyzer
calibrations
Calibration History - in the form of ASCII data les that
can be opened with Microsoft “Notepad” as simple .txt le
formats. The content can then be copied and pasted into
Microsoft Excel spread. Each Calibration and/or Validation
event is logged with the date (MM-DD-YYYY) and time
that it
occurred, the Mode (Calibration or Validation), the specic
type (Transmission Calibration, Zero Calibration, etc.) and
the ‘K’ (constant) factor used for the event
Example:
11-14-2007 14:11:19 span_calibrate 612814.89
9250.84 152665.88
11-14-2007 14:50:16 transmission_cal 14.95
calibr .bak
Calibration History
Back-Up
A back-up of historical log
of analyzer calibrations
Calibration History - When the calibr.his le exceeds
100KB size the contents is saved to this .bak le and the
.his le is emptied. This .bak le is in the form of ASCII
data les that can be opened with Microsoft “Notepad” as
simple .txt le
formats. The content can then be copied and pasted into
Microsoft Excel spread. Each Calibration and/or Validation
event is logged with the date (MM-DD-YYYY) and time
that it occurred, the Mode (Calibration or Validation), the
specic
type (Transmission Calibration, Zero Calibration, etc.) and
the ‘K’ (constant) factor used for the event
Example:
11-14-2007 14:11:19 span_calibrate 612814.89
9250.84 152665.88
11-14-2007 14:50:16 transmission_cal 14.95