<8. VALIDATION AND CALIBRATION>
• Check Gas Pressure species the pressure at which the gas within the online line check gas ow cell.
Typically this cell is vented to atmosphere so an atmospheric pressure value (14.7 psiA or 1.01barA) will work.
• Valve Selection species which analyzer’s solenoid valve driver is used for the check gas.
• Check Gas Purge Time species how long the check gas will purge the check gas ow cell.
Normal gas Purge Time species how long N
2
will purge the check gas ow cell.
• Validation Reading species the type of reporting method for the validation concentration (see earlier
explanation of these types).
• Analog Output Mode species Block, Track, or Hold of all 4 to 20 mA output during online check.