<8. VALIDATION AND CALIBRATION>
• Check Gas Pressure species the pressure at which the gas within the online line check gas ow cell.
Typically this cell is vented to atmosphere so an atmospheric pressure value (14.7 psiA or 1.01barA) will work.
• Valve Selection species which analyzer’s solenoid valve driver is used for the check gas.
• Check Gas Purge Time species how long the check gas will purge the check gas ow cell.
• Normal gas Purge Time species how long N
2
will purge the check gas ow cell.
• Validation Reading species the type of reporting method for the validation concentration (see earlier
explanation of these types).
• Analog Output Mode species Block, Track, or Hold of all 4 to 20 mA output during online check.